BRIEF SCOPE AND TECHNICAL SPECIFICATIONS OF “PROCUREMENT OF C&I

LAB INSTRUMENTS”

The scope of the subject procurement is “C&I Lab Instruments”, as per the technical specifications given below:

Item No 1:

Electronic Test Bench:

Selfcontained test bench made of anodized aluminium for testing and repair of electronic instruments and gadgets shall be fur- nished. It shall be modular in design enabling easy change of ar- rangement. The test bench shall have approximate dimension (FREE SPACE IN FRONT) (LxWxH) 1000x500x750 (dimension in mm). Test bench shall be furnished with thick laminated top. The front panel shall be made from 3-mm thick anodized aluminum plate and individually engraved in mimic style.

Bench-Top Unit : It should have Built-in modules with suitable cutout to equip / mount various instruments. The main power pre- wired to cater all the instruments mounted with frontal controls in this bench-top unit.

Under Bench Cabinet : It should have Storage space with lockable drawers (3 Nos.) for keeping necessary stationary items / tools and tackles, etc The test bench includes socket holders. The front equipment should include multi point connectors and test jacks. The test bench shall include the following module.

1. Main Power supply module

240 VAC, 50 Hz single phase main power supply

Residual current circuit breaker

Power Socket : 3 Nos. of 240V AC / 5A standard Indian make sock- ets

2.Servo Controlled Voltage Stabilizer of 240 V AC ±0.5%, 1 Phase,

50Hz, 3kVA Reputed Indian Make Specifications

Supply voltage : 240V AC, 50 Hz

Output voltage : 220V AC ± 5%, 50Hz / 3kVA

Stability : within ± 0.5% servo – controlled stabilizer Power Socket : 6 Nos. of 240V AC / 5A

Power Socket : 1 No. of 110V AC / 10A standard Indian make sockets

3.Variable AC Power Supply - Auto Transformer (Variac)

Input Volts : 240 VAC, 50Hz, 1 Phase Output Volts : 0 to 270 V AC

Current rating : 10 A

Digital Meter (V & I) : 0 to 300V AC - 1 No.

0 to 10 A - 1 No. Digital Ammeter & Voltmeter of 3 1/2 digit with resolution of 0.001 to 1 count for 3 1/2 depending on the range

Power Socket : 240VAC /10A socket (Std. Indian Make) - 1 No. with On-Off lamp indication switch.

4.Dual Variable regulated (0-64V / 0-5A) DC Power Supply

• Digital output metering 3 digit DPM

• High stability and close regulation ± 0.01% Specifications Dual Range 0 to 64 VDC / 0 to 5 Amps

Metering 3 Digit DPM Meter accuracy ± 3 counts Regulation :

Line ± 0.01% ± 2mv for ± 10% change in line voltage

Load ± 0.01% ± 2mv for load change from zero to full load

Power Socket : 24VDC socket (Std. Indian Make) - 1 No. with On-Off lamp indication switch.

5.Fixed DC Power Supply

Input Voltage : 240 VAC, ± 10%, 1f, 50Hz

Output Voltage & Current :

1)±12VDC / 10A - 1 No

2)±24VDC / 10A - 1 No.

3)±48 VDC / 5A - 1 No.

Display : Digital Voltmeter & Ammeter (for each fixed DC Power Supply)

No. of Sockets : 2 sets of O/P sockets will be provided for each power supply

6.Digital Multimeter(6 1/2 Digit) Accuracy 0.1% rdg

7.Test Leads

Make : Standard Indian Make

Qty. : 5 Sets.

8.Standard High Accuracy Resistors Standard Resistance of capacity 0.25W, Accuracy : 0.1%, of resistance of 1 OHM, 10

OHM, 25 OHM, 50 OHM, 100 OHM, 250 OHM, (1 Set )

9.DC voltage stabilizer DC power supply of voltage 0-64VDC/0-5A of regulation of ± 0.01% under facility

Necessary control devices for voltage and current adjust- ment/control, switches for power supply to feeders, isolation etc., shall be included. All electric circuits shall be protected

against overload by fuses, whose holders are located on front of the panel. All electrical circuits and metal structure shall be properly grounded. 2 nos. of antistatic rubber mat and 4 nos. wrist band shall be provided for working on PCB , electronic cards etc

Installation will be in vendor scope, 10 % amount will be retained till installation of bench.

detailed o&m manual and trainning shall be provided by supplier

_____________________________________________________________________________

Item No 2:

TECHNICAL SPECIFICATION FOR PORTABLE DIGITAL TERMOCOUPLE & RTD SIMULATOR CUM MEASUREMENT.

1.TYPE: SOURCE& MEASUREMET OF T/C,RTD,mV & OHMS

2.DISPLAY :4 ½ DIGITt LCD DISPLAY .

3.POWER SOURCE: 4 x A A SIZE BATTERIERS/OTHER SUITABLE BATTERIES WITH

230 V , 50 Hz AC ADAPTOR AS ALTERNATE POWER SUPPLY.

4.AUTO POWER OFF

5.TEMPERATURE COMPENSTION : INTERNAL REFERRENCE JUNCTION COMPEN- SATION

6.SOURCE & MEASUREMENT : SUITABLE FOR ALL TYPES OF RTD AND THERMO- COUPLE MEASUREMENT AND SIMULATION

i)K TYPE T/C : RANGE: (-) 200 - 1200 DEG C

 

 

RESOLUTION

 

:

0.1

DEG C

ii) E TYPE T/C :

RANGE: (-)

200

- 1000 DEG C

 

 

RESOLUTION

 

:

0.1

DEG C

iii)J TYPE T/C :

RANGE: (-)

200

- 1200 DEG C

 

 

RESOLUTION

 

:

0.1

DEG C

iv)T TYPE T/C :

RANGE: (-)

200

- 400 DEG C

 

 

RESOLUTION

 

:

0.1

DEG C

v)R TYPE T/C:

 

RANGE:

0

-

1600 DEG C

 

 

RESOLUTION

 

:

0.1

DEG C

vi)S TYPE T/C:

 

RANGE:

0

- 1600

DEG

C

 

 

RESOLUTION

 

:

0.1

DEG C

vii)B TYPE T/C:

 

RANGE:

600

-

1800 DEG C

 

 

RESOLUTION

 

:

0.1

DEG C

xi)RTD Pt 100 :

RANGE (-) 200 - 850 DEG C

 

 

RESOLUTION

 

:

0.1

DEG C

xii)400 OHMS

:

RANGE

0

- 400 OHMS

 

 

 

RESOLUTION

 

:

0.1

OHMS

xiii)100 mV

:

RANGE

0

- / + 100 mV

 

 

RESOLUTION

 

:

10 MICRO V

7) ACCURACY OF SOURCE

:

0.05

%

OR BETTER

8) ACCURACY OF MEASUREMENT:

0.07

%

OR BETTER

9)ACCESSORIES WITH EACH CALIBRATOR:

i)Instruction Manuals,

ii)Carrying case,

iii)Connection Leads

iv)Set of Batteries.

v)Standard K - Type & R-Type T/C probes , 1000mm lenght

vi)Standard RTD Pt - 100 Probe , 1000 mm Length

10. Calibration ;

The calibration of the equipment shall be

 

traceable to National/International Stand-

ards.

 

 

 

Calibration certificate traceable

to

National/International Standards. is to be supplied along with the equipments

detailed o&m manual and trainning shall be provided by supplier

________________________________________________________________________

Item No 3:

TECHNICAL SPECIFICATION FOR VACCUME CALIBRATOR Pressure Range -1 kg (vaccume) to 1 kg

potable vaccume pressure calibrator for field use Internal battery powered

INTERNAL vac/pr source DRAING INTERNAL VENT VALVE

RISING /FALLING FINE ADJUSTMENT VALVE ACCURACY 0.05% FS or better RESOLUTION 1 MiliBAR

ENCLOUSER IP-54 DISPLAY BACKLIGHT LCD

UNITS BAR,PASCAL,KG,PSI,MMWCL OPERATING TEMP UPTO 40 DEG

ACESSORIES TUBES,FITTINGS FOR ALL MALE FEMALE CONNECTORS & CARRI- ING CASE, TEST LEADS,charger for recharabeble battery

DOCUMENTS USER GUIDE,

Calibration ; The calibration of the equipment shall be traceable to National/International Stand-

ards.

Calibration certificate traceable to National/International Standards. is to be supplied along with the equipments

________________________________________________________________________

Item No 4:

Electrically operated vaccum pump.An electrically operated vacuum pump complete with vacuum gauges (0-760 mm Hg), lub oil level switches, moisture trap, inlet dust filter, exhaust mist filter etc. as necessary for the pump shall be provided. The vacuum pump shall be single stage, oil sealed, rotary vane type offering high degree of efficiency combined with minimum maintenance. Vacuum pump shall be capable of providing very high degree of vacuum typically 3 x 10-2 mbar or better. The pump shall have high water vapour tolerance typically greater than 15 mbar. The pump shall have its own arrangement for air cooling if required lub oil level switches, moisture trap, inlet dust filter, exhaust mist filter, etc. as necessary for the pump shall be included.

detailed o&m manual and trainning shall be provided by supplier

________________________________________________________________________

Item No 5:

U TUBE MANOMETER

Range : (-) 1000-0-1000 mm Minimum Division: 1 mm U-Tube type primary standard manometer

Stand having levelling knob and attached spirit level bulb detailed o&m manual and trainning shall be provided by supplier

________________________________________________________________________

Item No 6:

Sensor Type : S Type Platinum/Platinum-Rhodium THERMOCOUPLE Minimum Diameter :7 mm

Minimum Length :600 mm Temperature range : 0-1600 °C

Uncertainty : 0.8 ° C ( 2.2 ° C for higher temperature )

S TYPE STANDARD THERMOCOUPLE WILL BE USED FOR STANDARDIZE THE MEASUREMENT

Calibration Certificate Includes 4 point UKAS Calibration up to 1100°C (Calibration Certificate). The calibration of the equipment shall be

traceable to National/International Standards. Calibration certif- icate traceable to National/International Standards. is to be sup- plied along with the equipments

Accessories Carrying case, Connecting Cable, Extension Cable detailed o&m manual and trainning shall be provided by supplier

________________________________________________________________________

Item No 7:

Pneumatic test bench for testing pnematic transmitters, control- lers, power cylinders,control valves etc

Construction details of Work Bench : Pneumatic test bench should be made up of anodized aluminum & modular in design, enabling easy change of layout / arrangement, sturdy steel frame & sleek laminated top table, space for storing fittings / thermo wells. Space for customer's test equipment, customer's add-ons, modular construction - easy servicing. The approximate pneumatic test bench dimension (FRONT FREE WORKING SPACE)will be 1400 x 750 x 750 mm (Lx W x H)

Bench-top unit With built-in modules, blank plate with suitable cutout to equip / mount various instruments. The mains power pre- wired to cater all the instruments mounted with frontal controls in this bench-top unit. Under bench cabinet Storage space with lockable drawers for keeping necessary stationary items / tools and tackles, etc.

Bench should have following modular facility Facility : 1

Mains power supply Make : Standard Indian Make * 230 VAC, 50 Hz single phase main power supply * For protection of built-in mod- ules/cassettes etc. * Residual current circuit breaker * Rated fault current : 30mA * Rated current : 25A * No. of poles : 2 * Emergency off * Line indicating lamp

Facility 2

Air Supply Outlet: 0-20 psi and 0-30 psi with Two nos. 4 1/2 digit pressure indicators of range vacuum to 20000 mm Wc and 0-30 psi, accuracy 0.05% of full scale shall be provided.

Facility : 3

air supply outlet facility with Output quick release coupling, Air filter and automatic pressure controller/calibrator and digital pressure indicator for

1.0-2 kg/cm2

2.0-4 kg/cm2

3.0-7 kg/cm2

Communication with PC for calibration record management Facility :4

Dual 0-32V DC Regulated Power Supply

Features • Digital output metering 3 digit DPM • Constant voltage / constant current operation • High stability and close regulation

± 0.01% • Proven reliability and endurance

Facility : 5

Digital MilliAmp / Millivolt Calibrator (source)

Features : • Accuracy ±0.02% rdg, Dual channels • Large graphic backlight display • Communication with PC for calibration record management • mV, V, mA (active & passive loop) • Four operative mode : measure, simulate, measure/simulate and measure/measure •

Measure and simulation of Active & Passive 4-20mA current loop • Measure and simulation of DC Voltage : range 0-200mV, 0-2V and 0- 20V Specifications : 0-50mA current source & 0-200mV voltage source

Facility :6

Accessories • Standard connectors: quick connectors with PVC hose.

• Adaptors 1/4" BSP Male x 1/8" BSP Parallel Female – 1 No. 1/4" BSP Male x 1/8" NPT Female – 2 Nos. 1/4" BSP Male x 1/4" NPT Fe- male – 2 Nos. 1/4" BSP Male x 3/8" BSP Parallel Female – 1 No.

1/4" BSP Male x 3/8" NPT Female – 1 No. 1/4" BSP Male x 1/2" BSP Parallel Female – 1 No. 1/4" BSP Male x 1/2" NPT Female – 1 No. • Tools (spanners) : double end spanner from 6 - 7 to 32. Each of the Facility is one No.

Installation will be in vendor scope, 10 % amount will be retained till installation of bench.

Calibration

The calibration of the equipment shall be

 

 

traceable to National/International Stand-

ards.

 

 

 

Calibration certificate traceable

to

 

National/International Standards. is to be

 

supplied along with the equipments

 

detailed o&m manual,software for calibration record management and trainning shall be provided by supplier

________________________________________________________________________

Item No 8:

DIGITAL MASTER PRESSURE GAUGE.

1.RANGE: 0 TO 500 KSC

2.OVER RANGE PROTECTION 130% OF FULL SCALE

3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING NON-LINEARITY, REPEATABILITY, HYSTERESIS AND TEMPERATURE EFFECTS FROM 0 TO 50 DEG C.

4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER

5.LCD WITH BACKLIGHT DISPLAY

6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.

7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.

8.PROCESS CONNECTION: 1/2" NPT MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 NPT TO 1/2 BSP.

9.OPERATING TEMPERATURE: 0-50 DEG C.

10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.

11.SENSOR MATERIAL: SS-316

12.WELDED PARTS: SS316

13.ENCLOSURE: IP 65 PROTECTION CLASS.

14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.

15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.

16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.

17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.

THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.

________________________________________________________________________

Item No 9:

Technical Specifications for Digital Storage Oscilloscope

1.Analog Bandwidth : 100MHz or higher

2.Input Channels : Minimum 2 Channels

3.Real time sample rate :1 GS/sec

4.Equivalent time sample rate : 2.5 TS/sec or better

5.Input Sensitivity : 2mV/div to 5V/div

6.Input Coupling : AC, DC, Ground DC50 ohm

7.Input Impedance :1Mohm, 50ohm

8.Vertical Resolution : 8bits (13 bits in High res mode) or high-

er

9.DC gain accuracy : ± 1.5 % or better

10.Time base range : 500ps/div to 50s/div 11.Record length : 2.5MW/ch or more 12.Interpolation: Sin x/x interpolation 13.Acquisition modes :25,000 wfms /sec/Ch

14.Main trigger modes : Auto, Auto level, Normal, Single and N Single.

15.Acquisition rate in N Single :2.5 Million waveform/Sec.

16.Trigger types: Edge/state, width, event interval, serial pat- tern,

window triggering.

17.Automatic waveform measurements : 16 automatic measuring param- eter

displayed at a time like: Min, Max, High, Low, P-P, High-Low, +Over,

-Over, RMS, Mean, Sdev, IntegTY,C.rms, C.mean, C.Sdev, C.IntegTY,

(1/Freq),Freq, Count,Burst, +Width, -Width, Period, Duty, Rise, Fall, Delay etc.

18.Waveform maths : Add, Subtract, Divide, Integration, Count (Edge), Count (Rotary), Through, Delay, Moving Avg, Low Pass, High Pass.

Computation of up to 8 traces .

19.Analysis Function: Selectable from XY, FFT, Wave parameter, Histogram and Serial Bus.

20.Waveform storage : Internet Flash ROM 90MB.USB peripheral sup- port, PC card interface(Flash ATA memory card),Internal HDD of minimum 32 GB.

21.I/O PORTS :USB communication , GPIB interface, Ethernet Inter- face.

22.Software : To View & analyze the data on PC.

23.Additional features :History memory to view and analyze up to minimum 1000 previously acquired waveforms, history reply function, dual zoom facility, Security for confidential tests, advance dis- play technology (Dot Density Display) Mask Testing, Histogram dis- plays, Statistics, Calculations, Real-Time Analog/Digital Filter- ing, Trend Displays etc.

24.Current and Differential Probes to be supplied: a)Current Probes : 02 nos.

b)Differentail Probes : DC to 100 MHZ, +/- 700V-- 02 Nos.

25.Standard accessories :Power cable – 01 no, Passive probes (4 nos.),

User's Manual (2 sets), any other accessories required to function oscilloscope and for pc connectivity

26.Weight : Not more than 10 kgclip, 4 mm add-on probe tip, ground lead to 4 mm banana plug

________________________________________________________________________

Item No 10:

Sensor Type : Platinum resistance element, Four wire type Outer Sheath material Quartz/SS/INCONEL

Minimum Diameter : 7.5mm Minimum Length : 500 mm Stability : Better than 0.03 °C Temperature range : 0-600 °C

RTD WILL BE USED FOR STANDARDIZE THE MEASUREMENT Accessories

Carrying case shall be supplied - The sensor has a cable of 2mtr. (Connecting Cable, Extension Cable)

Calibration ; The calibration of the equipment shall be traceable to National/International Standards.Calibration certifi-

cate traceable

to

National/International Standards. is to be

supplied along

with

the equipments .

Detailed o&m manual and trainning shall be provided by supplier

________________________________________________________________________

Item No 11:

Digital TEMPERATURE BATH SHOULD MEET FOLLOWING REQUIRTEMENTS Temperature range : 50-700 Deg C

Type : Microprocessor based programming with auto calibration fa- cility comprising of air fitter, air pressure regulator Stability : ± 0.01 °C (at 550 °C) or better over period of 20 minutes

Uniformity: +/-0.02°C (at 550 °C) or better Power supply : 240VAC, 50 Hz -

PC interface : RS 232 with windows driven software with capacity to control, automatically log and display temp of bath with graph- ical presentation

Features : Protection Fuses, overheating protection with low flow cutout, over pressure, alarms for low flow, blocked filter (with auto cutout) etc designed to eliminate powder loss

Accessories : Equalizing block, PC communication interface soft- ware and manual(operating) AND ALL REQUIRED ACCESSORIES NEEDED TO FUNCTION

Standards Designed : as per latest IEC safety standards

Calibration ; The calibration of the equipment shall be traceable to National/International Stand-

ards.

Calibration certificate traceable to National/International Standards. is to be supplied along with the equipments

detailed o&m manual and trainning shall be provided by supplier

________________________________________________________________________

Item No 12:

DIRECT MOUNTING TYPE DIGITAL PRESSURE CALIBRATOR (WITH INBUILT PRESSURE SENSOR) SUITABLE TO HYDRAULIC COMPARATOR.RANGE:0 TO 20 KG/SQCM. SPECS : PRESSURE RANGE: 0 TO 20 KG/CM2 OR BAR PRESSURE UNIT: BAR,MMWC, MMHG,INHG,KG/CM2. VOLTAGE MEASUREMENT: 0 TO 10V DC,ACCURACY: +/-0.10%FSD. CURRENT MEASUREMENT: 0 TO 24 MA DC,ACCURACY: +/- 0.1%FSD. POWER SUPPLY:RECHARGABLE BAT- TERY/MAINS(230V AC,50HZ) O PERATION VIA CHARGER OPERATING TEMPERA- TURE: 0 TO 40 DEG C. PRESSURE MEDIA: AIR/OIL ZERO ADJUSTMENTS: AU- TO ZERO OR MANUAL KEYBOARD OPERATION. RESOLUTION: MINIMUM 5 DIGITS VOLTAGE OUTPUT:PROVISION OF 24V DC SUPPLY FOR PRESSURE TRANSMITTER. PRESSURE CON NECTION:1/2"NPT MALE. ACCURACY:0.025% FSD. ACCESSO- RIES:POWER CORD,BATTERY CHARGER,CONNECTING LEADS,CLIPS ETC. CALI- BRATION CERTIFICATES FROM NABL/INTERNATIONAL ACCREDITED LABORATORY

________________________________________________________________________

Item No 13:

DIGITAL MASTER PRESSURE GAUGE.

1.RANGE: 0 TO 250 KSC

2.OVER RANGE PROTECTION 130% OF FULL SCALE

3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING NON-LINEARITY, REPEATABILITY, HYSTERESIS AND TEMPERATURE EFFECTS FROM 0 TO 50 DEG C.

4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER

5.LCD WITH BACKLIGHT DISPLAY

6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.

7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.

8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION.

9.OPERATING TEMPERATURE: 0-50 DEG C.

10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.

11.SENSOR MATERIAL: SS-316

12.WELDED PARTS: SS316

13.ENCLOSURE: IP 65 PROTECTION CLASS.

14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.

15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.

16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.

17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.

THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.

________________________________________________________________________

Item No 14:

DIGITAL MASTER PRESSURE GAUGE.

1.RANGE: 0 TO 100 BAR

2.OVER RANGE PROTECTION 130% OF FULL SCALE

3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING NON-LINEARITY, REPEATABILITY, HYSTERESIS AND TEMPERATURE EFFECTS FROM 0 TO 50 DEG C.

4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER

5.LCD WITH BACKLIGHT DISPLAY

6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.

7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.

8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION.

9.OPERATING TEMPERATURE: 0-50 DEG C.

10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.

11.SENSOR MATERIAL: SS-316

12.WELDED PARTS: SS316

13.ENCLOSURE: IP 65 PROTECTION CLASS.

14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.

15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.

16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.

17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.

THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.

________________________________________________________________________

Item No 15:

DIGITAL MASTER PRESSURE GAUGE.

1.RANGE: (-1) TO 0 BAR

2.OVER RANGE PROTECTION 130% OF FULL SCALE

3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING NON-LINEARITY, REPEATABILITY, HYSTERESIS AND TEMPERATURE EFFECTS FROM 0 TO 50 DEG C.

4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER

5.LCD WITH BACKLIGHT DISPLAY

6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.

7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.

8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION

9.OPERATING TEMPERATURE: 0-50 DEG C.

10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.

11.SENSOR MATERIAL: SS-316

12.WELDED PARTS: SS316

13.ENCLOSURE: IP 65 PROTECTION CLASS.

14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.

15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.

16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.

17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.

THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.

________________________________________________________________________

Item No 16:

DIGITAL MASTER PRESSURE GAUGE.

1.RANGE: 0 TO 50 BAR

2.OVER RANGE PROTECTION 130% OF FULL SCALE

3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING NON-LINEARITY, REPEATABILITY, HYSTERESIS AND TEMPERATURE EFFECTS FROM 0 TO 50 DEG C.

4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER

5.LCD WITH BACKLIGHT DISPLAY

6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.

7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.

8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION9. OPERATING TEMPERATURE: 0-50 DEG C.

10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.

11.SENSOR MATERIAL: SS-316

12.WELDED PARTS: SS316

13.ENCLOSURE: IP 65 PROTECTION CLASS.

14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.

15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.

16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.

17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.

THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.

________________________________________________________________________

Item No 17:

DIGITAL MASTER PRESSURE GAUGE.

1.RANGE: 0 TO 200 BAR

2.OVER RANGE PROTECTION 130% OF FULL SCALE

3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING NON-LINEARITY, REPEATABILITY, HYSTERESIS AND TEMPERATURE EFFECTS FROM 0 TO 50 DEG C.

4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER

5.LCD WITH BACKLIGHT DISPLAY

6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.

7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.

8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION

9.OPERATING TEMPERATURE: 0-50 DEG C.

10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION-

ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.

11.SENSOR MATERIAL: SS-316

12.WELDED PARTS: SS316

13.ENCLOSURE: IP 65 PROTECTION CLASS.

14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.

15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.

16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.

17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.

THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.

________________________________________________________________________

Item No 18:

DIGITAL MASTER PRESSURE GAUGE.

1.RANGE: 0 TO 4 KSC

2.OVER RANGE PROTECTION 130% OF FULL SCALE

3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING NON-LINEARITY, REPEATABILITY, HYSTERESIS AND TEMPERATURE EFFECTS FROM 0 TO 50 DEG C.

4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER

5.LCD WITH BACKLIGHT DISPLAY

6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.

7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.

8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION

9.OPERATING TEMPERATURE: 0-50 DEG C.

10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.

11.SENSOR MATERIAL: SS-316

12.WELDED PARTS: SS316

13.ENCLOSURE: IP 65 PROTECTION CLASS.

14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.

15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.

16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.

17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.

THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.

________________________________________________________________________

Item No 19:

DIGITAL MASTER PRESSURE GAUGE.

1.RANGE: (-1) TO 2 BAR

2.OVER RANGE PROTECTION 130% OF FULL SCALE

3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING NON-LINEARITY, REPEATABILITY, HYSTERESIS AND TEMPERATURE EFFECTS FROM 0 TO 50 DEG C.

4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER

5.LCD WITH BACKLIGHT DISPLAY

6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.

7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.

8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION

9.OPERATING TEMPERATURE: 0-50 DEG C.

10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.

11.SENSOR MATERIAL: SS-316

12.WELDED PARTS: SS316

13.ENCLOSURE: IP 65 PROTECTION CLASS.

14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.

15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.

16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.

17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.

THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.