BRIEF SCOPE AND TECHNICAL SPECIFICATIONS OF “PROCUREMENT OF C&I
LAB INSTRUMENTS”
The scope of the subject procurement is “C&I Lab Instruments”, as per the technical specifications given below:
Item No 1:
Electronic Test Bench:
Selfcontained test bench made of anodized aluminium for testing and repair of electronic instruments and gadgets shall be fur- nished. It shall be modular in design enabling easy change of ar- rangement. The test bench shall have approximate dimension (FREE SPACE IN FRONT) (LxWxH) 1000x500x750 (dimension in mm). Test bench shall be furnished with thick laminated top. The front panel shall be made from
Under Bench Cabinet : It should have Storage space with lockable drawers (3 Nos.) for keeping necessary stationary items / tools and tackles, etc The test bench includes socket holders. The front equipment should include multi point connectors and test jacks. The test bench shall include the following module.
1. Main Power supply module
•240 VAC, 50 Hz single phase main power supply
•Residual current circuit breaker
•Power Socket : 3 Nos. of 240V AC / 5A standard Indian make sock- ets
2.Servo Controlled Voltage Stabilizer of 240 V AC ±0.5%, 1 Phase,
50Hz, 3kVA Reputed Indian Make Specifications
Supply voltage : 240V AC, 50 Hz
Output voltage : 220V AC ± 5%, 50Hz / 3kVA
Stability : within ± 0.5% servo – controlled stabilizer Power Socket : 6 Nos. of 240V AC / 5A
Power Socket : 1 No. of 110V AC / 10A standard Indian make sockets
3.Variable AC Power Supply - Auto Transformer (Variac)
Input Volts : 240 VAC, 50Hz, 1 Phase Output Volts : 0 to 270 V AC
Current rating : 10 A
Digital Meter (V & I) : 0 to 300V AC - 1 No.
0 to 10 A - 1 No. Digital Ammeter & Voltmeter of 3 1/2 digit with resolution of 0.001 to 1 count for 3 1/2 depending on the range
Power Socket : 240VAC /10A socket (Std. Indian Make) - 1 No. with
4.Dual Variable regulated
• Digital output metering 3 digit DPM
• High stability and close regulation ± 0.01% Specifications Dual Range 0 to 64 VDC / 0 to 5 Amps
Metering 3 Digit DPM Meter accuracy ± 3 counts Regulation :
Line ± 0.01% ± 2mv for ± 10% change in line voltage
Load ± 0.01% ± 2mv for load change from zero to full load
Power Socket : 24VDC socket (Std. Indian Make) - 1 No. with
5.Fixed DC Power Supply
Input Voltage : 240 VAC, ± 10%, 1f, 50Hz
Output Voltage & Current :
1)±12VDC / 10A - 1 No
2)±24VDC / 10A - 1 No.
3)±48 VDC / 5A - 1 No.
Display : Digital Voltmeter & Ammeter (for each fixed DC Power Supply)
No. of Sockets : 2 sets of O/P sockets will be provided for each power supply
6.Digital Multimeter(6 1/2 Digit) Accuracy 0.1% rdg
7.Test Leads
Make : Standard Indian Make
Qty. : 5 Sets.
8.Standard High Accuracy Resistors Standard Resistance of capacity 0.25W, Accuracy : 0.1%, of resistance of 1 OHM, 10
OHM, 25 OHM, 50 OHM, 100 OHM, 250 OHM, (1 Set )
9.DC voltage stabilizer DC power supply of voltage
Necessary control devices for voltage and current adjust- ment/control, switches for power supply to feeders, isolation etc., shall be included. All electric circuits shall be protected
against overload by fuses, whose holders are located on front of the panel. All electrical circuits and metal structure shall be properly grounded. 2 nos. of antistatic rubber mat and 4 nos. wrist band shall be provided for working on PCB , electronic cards etc
Installation will be in vendor scope, 10 % amount will be retained till installation of bench.
detailed o&m manual and trainning shall be provided by supplier
_____________________________________________________________________________
Item No 2:
TECHNICAL SPECIFICATION FOR PORTABLE DIGITAL TERMOCOUPLE & RTD SIMULATOR CUM MEASUREMENT.
1.TYPE: SOURCE& MEASUREMET OF T/C,RTD,mV & OHMS
2.DISPLAY :4 ½ DIGITt LCD DISPLAY .
3.POWER SOURCE: 4 x A A SIZE BATTERIERS/OTHER SUITABLE BATTERIES WITH
230 V , 50 Hz AC ADAPTOR AS ALTERNATE POWER SUPPLY.
4.AUTO POWER OFF
5.TEMPERATURE COMPENSTION : INTERNAL REFERRENCE JUNCTION COMPEN- SATION
6.SOURCE & MEASUREMENT : SUITABLE FOR ALL TYPES OF RTD AND THERMO- COUPLE MEASUREMENT AND SIMULATION
i)K TYPE T/C : RANGE:
|
|
RESOLUTION |
|
: |
0.1 |
DEG C |
||
ii) E TYPE T/C : |
RANGE: |
200 |
- 1000 DEG C |
|||||
|
|
RESOLUTION |
|
: |
0.1 |
DEG C |
||
iii)J TYPE T/C : |
RANGE: |
200 |
- 1200 DEG C |
|||||
|
|
RESOLUTION |
|
: |
0.1 |
DEG C |
||
iv)T TYPE T/C : |
RANGE: |
200 |
- 400 DEG C |
|||||
|
|
RESOLUTION |
|
: |
0.1 |
DEG C |
||
v)R TYPE T/C: |
|
RANGE: |
0 |
- |
1600 DEG C |
|||
|
|
RESOLUTION |
|
: |
0.1 |
DEG C |
||
vi)S TYPE T/C: |
|
RANGE: |
0 |
- 1600 |
DEG |
C |
||
|
|
RESOLUTION |
|
: |
0.1 |
DEG C |
||
vii)B TYPE T/C: |
|
RANGE: |
600 |
- |
1800 DEG C |
|||
|
|
RESOLUTION |
|
: |
0.1 |
DEG C |
||
xi)RTD Pt 100 : |
RANGE |
|||||||
|
|
RESOLUTION |
|
: |
0.1 |
DEG C |
||
xii)400 OHMS |
: |
RANGE |
0 |
- 400 OHMS |
|
|||
|
|
RESOLUTION |
|
: |
0.1 |
OHMS |
||
xiii)100 mV |
: |
RANGE |
0 |
|||||
|
|
RESOLUTION |
|
: |
10 MICRO V |
|||
7) ACCURACY OF SOURCE |
: |
0.05 |
% |
OR BETTER |
||||
8) ACCURACY OF MEASUREMENT: |
0.07 |
% |
OR BETTER |
9)ACCESSORIES WITH EACH CALIBRATOR:
i)Instruction Manuals,
ii)Carrying case,
iii)Connection Leads
iv)Set of Batteries.
v)Standard K - Type &
vi)Standard RTD Pt - 100 Probe , 1000 mm Length
10. Calibration ; |
The calibration of the equipment shall be |
|
|
traceable to National/International Stand- |
|
ards. |
|
|
|
Calibration certificate traceable |
to |
National/International Standards. is to be supplied along with the equipments
detailed o&m manual and trainning shall be provided by supplier
________________________________________________________________________
Item No 3:
TECHNICAL SPECIFICATION FOR VACCUME CALIBRATOR Pressure Range
potable vaccume pressure calibrator for field use Internal battery powered
INTERNAL vac/pr source DRAING INTERNAL VENT VALVE
RISING /FALLING FINE ADJUSTMENT VALVE ACCURACY 0.05% FS or better RESOLUTION 1 MiliBAR
ENCLOUSER
UNITS BAR,PASCAL,KG,PSI,MMWCL OPERATING TEMP UPTO 40 DEG
ACESSORIES TUBES,FITTINGS FOR ALL MALE FEMALE CONNECTORS & CARRI- ING CASE, TEST LEADS,charger for recharabeble battery
DOCUMENTS USER GUIDE,
Calibration ; The calibration of the equipment shall be traceable to National/International Stand-
ards.
Calibration certificate traceable to National/International Standards. is to be supplied along with the equipments
________________________________________________________________________
Item No 4:
Electrically operated vaccum pump.An electrically operated vacuum pump complete with vacuum gauges
detailed o&m manual and trainning shall be provided by supplier
________________________________________________________________________
Item No 5:
U TUBE MANOMETER
Range :
Stand having levelling knob and attached spirit level bulb detailed o&m manual and trainning shall be provided by supplier
________________________________________________________________________
Item No 6:
Sensor Type : S Type
Minimum Length :600 mm Temperature range :
Uncertainty : 0.8 ° C ( 2.2 ° C for higher temperature )
S TYPE STANDARD THERMOCOUPLE WILL BE USED FOR STANDARDIZE THE MEASUREMENT
Calibration Certificate Includes 4 point UKAS Calibration up to 1100°C (Calibration Certificate). The calibration of the equipment shall be
traceable to National/International Standards. Calibration certif- icate traceable to National/International Standards. is to be sup- plied along with the equipments
Accessories Carrying case, Connecting Cable, Extension Cable detailed o&m manual and trainning shall be provided by supplier
________________________________________________________________________
Item No 7:
Pneumatic test bench for testing pnematic transmitters, control- lers, power cylinders,control valves etc
Construction details of Work Bench : Pneumatic test bench should be made up of anodized aluminum & modular in design, enabling easy change of layout / arrangement, sturdy steel frame & sleek laminated top table, space for storing fittings / thermo wells. Space for customer's test equipment, customer's
Bench should have following modular facility Facility : 1
Mains power supply Make : Standard Indian Make * 230 VAC, 50 Hz single phase main power supply * For protection of
Facility 2
Air Supply Outlet:
Facility : 3
air supply outlet facility with Output quick release coupling, Air filter and automatic pressure controller/calibrator and digital pressure indicator for
1.
2.
3.
Communication with PC for calibration record management Facility :4
Dual
Features • Digital output metering 3 digit DPM • Constant voltage / constant current operation • High stability and close regulation
± 0.01% • Proven reliability and endurance
Facility : 5
Digital MilliAmp / Millivolt Calibrator (source)
Features : • Accuracy ±0.02% rdg, Dual channels • Large graphic backlight display • Communication with PC for calibration record management • mV, V, mA (active & passive loop) • Four operative mode : measure, simulate, measure/simulate and measure/measure •
Measure and simulation of Active & Passive
Facility :6
Accessories • Standard connectors: quick connectors with PVC hose.
• Adaptors 1/4" BSP Male x 1/8" BSP Parallel Female – 1 No. 1/4" BSP Male x 1/8" NPT Female – 2 Nos. 1/4" BSP Male x 1/4" NPT Fe- male – 2 Nos. 1/4" BSP Male x 3/8" BSP Parallel Female – 1 No.
1/4" BSP Male x 3/8" NPT Female – 1 No. 1/4" BSP Male x 1/2" BSP Parallel Female – 1 No. 1/4" BSP Male x 1/2" NPT Female – 1 No. • Tools (spanners) : double end spanner from 6 - 7 to 32. Each of the Facility is one No.
Installation will be in vendor scope, 10 % amount will be retained till installation of bench.
Calibration |
The calibration of the equipment shall be |
|
|
traceable to National/International Stand- |
|
ards. |
|
|
|
Calibration certificate traceable |
to |
|
National/International Standards. is to be |
|
|
supplied along with the equipments |
|
detailed o&m manual,software for calibration record management and trainning shall be provided by supplier
________________________________________________________________________
Item No 8:
DIGITAL MASTER PRESSURE GAUGE.
1.RANGE: 0 TO 500 KSC
2.OVER RANGE PROTECTION 130% OF FULL SCALE
3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING
4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER
5.LCD WITH BACKLIGHT DISPLAY
6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.
7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.
8.PROCESS CONNECTION: 1/2" NPT MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 NPT TO 1/2 BSP.
9.OPERATING TEMPERATURE:
10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.
11.SENSOR MATERIAL:
12.WELDED PARTS: SS316
13.ENCLOSURE: IP 65 PROTECTION CLASS.
14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.
15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.
16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.
17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.
THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.
________________________________________________________________________
Item No 9:
Technical Specifications for Digital Storage Oscilloscope
1.Analog Bandwidth : 100MHz or higher
2.Input Channels : Minimum 2 Channels
3.Real time sample rate :1 GS/sec
4.Equivalent time sample rate : 2.5 TS/sec or better
5.Input Sensitivity : 2mV/div to 5V/div
6.Input Coupling : AC, DC, Ground DC50 ohm
7.Input Impedance :1Mohm, 50ohm
8.Vertical Resolution : 8bits (13 bits in High res mode) or high-
er
9.DC gain accuracy : ± 1.5 % or better
10.Time base range : 500ps/div to 50s/div 11.Record length : 2.5MW/ch or more 12.Interpolation: Sin x/x interpolation 13.Acquisition modes :25,000 wfms /sec/Ch
14.Main trigger modes : Auto, Auto level, Normal, Single and N Single.
15.Acquisition rate in N Single :2.5 Million waveform/Sec.
16.Trigger types: Edge/state, width, event interval, serial pat- tern,
window triggering.
17.Automatic waveform measurements : 16 automatic measuring param- eter
displayed at a time like: Min, Max, High, Low,
(1/Freq),Freq, Count,Burst, +Width,
18.Waveform maths : Add, Subtract, Divide, Integration, Count (Edge), Count (Rotary), Through, Delay, Moving Avg, Low Pass, High Pass.
Computation of up to 8 traces .
19.Analysis Function: Selectable from XY, FFT, Wave parameter, Histogram and Serial Bus.
20.Waveform storage : Internet Flash ROM 90MB.USB peripheral sup- port, PC card interface(Flash ATA memory card),Internal HDD of minimum 32 GB.
21.I/O PORTS :USB communication , GPIB interface, Ethernet Inter- face.
22.Software : To View & analyze the data on PC.
23.Additional features :History memory to view and analyze up to minimum 1000 previously acquired waveforms, history reply function, dual zoom facility, Security for confidential tests, advance dis- play technology (Dot Density Display) Mask Testing, Histogram dis- plays, Statistics, Calculations,
24.Current and Differential Probes to be supplied: a)Current Probes : 02 nos.
b)Differentail Probes : DC to 100 MHZ, +/-
25.Standard accessories :Power cable – 01 no, Passive probes (4 nos.),
User's Manual (2 sets), any other accessories required to function oscilloscope and for pc connectivity
26.Weight : Not more than 10 kgclip, 4 mm
________________________________________________________________________
Item No 10:
Sensor Type : Platinum resistance element, Four wire type Outer Sheath material Quartz/SS/INCONEL
Minimum Diameter : 7.5mm Minimum Length : 500 mm Stability : Better than 0.03 °C Temperature range :
RTD WILL BE USED FOR STANDARDIZE THE MEASUREMENT Accessories
Carrying case shall be supplied - The sensor has a cable of 2mtr. (Connecting Cable, Extension Cable)
Calibration ; The calibration of the equipment shall be traceable to National/International Standards.Calibration certifi-
cate traceable |
to |
National/International Standards. is to be |
supplied along |
with |
the equipments . |
Detailed o&m manual and trainning shall be provided by supplier
________________________________________________________________________
Item No 11:
Digital TEMPERATURE BATH SHOULD MEET FOLLOWING REQUIRTEMENTS Temperature range :
Type : Microprocessor based programming with auto calibration fa- cility comprising of air fitter, air pressure regulator Stability : ± 0.01 °C (at 550 °C) or better over period of 20 minutes
Uniformity:
PC interface : RS 232 with windows driven software with capacity to control, automatically log and display temp of bath with graph- ical presentation
Features : Protection Fuses, overheating protection with low flow cutout, over pressure, alarms for low flow, blocked filter (with auto cutout) etc designed to eliminate powder loss
Accessories : Equalizing block, PC communication interface soft- ware and manual(operating) AND ALL REQUIRED ACCESSORIES NEEDED TO FUNCTION
Standards Designed : as per latest IEC safety standards
Calibration ; The calibration of the equipment shall be traceable to National/International Stand-
ards.
Calibration certificate traceable to National/International Standards. is to be supplied along with the equipments
detailed o&m manual and trainning shall be provided by supplier
________________________________________________________________________
Item No 12:
DIRECT MOUNTING TYPE DIGITAL PRESSURE CALIBRATOR (WITH INBUILT PRESSURE SENSOR) SUITABLE TO HYDRAULIC COMPARATOR.RANGE:0 TO 20 KG/SQCM. SPECS : PRESSURE RANGE: 0 TO 20 KG/CM2 OR BAR PRESSURE UNIT: BAR,MMWC, MMHG,INHG,KG/CM2. VOLTAGE MEASUREMENT: 0 TO 10V DC,ACCURACY:
________________________________________________________________________
Item No 13:
DIGITAL MASTER PRESSURE GAUGE.
1.RANGE: 0 TO 250 KSC
2.OVER RANGE PROTECTION 130% OF FULL SCALE
3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING
4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER
5.LCD WITH BACKLIGHT DISPLAY
6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.
7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.
8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION.
9.OPERATING TEMPERATURE:
10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.
11.SENSOR MATERIAL:
12.WELDED PARTS: SS316
13.ENCLOSURE: IP 65 PROTECTION CLASS.
14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.
15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.
16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.
17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.
THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.
________________________________________________________________________
Item No 14:
DIGITAL MASTER PRESSURE GAUGE.
1.RANGE: 0 TO 100 BAR
2.OVER RANGE PROTECTION 130% OF FULL SCALE
3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING
4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER
5.LCD WITH BACKLIGHT DISPLAY
6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.
7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.
8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION.
9.OPERATING TEMPERATURE:
10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.
11.SENSOR MATERIAL:
12.WELDED PARTS: SS316
13.ENCLOSURE: IP 65 PROTECTION CLASS.
14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.
15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.
16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.
17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.
THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.
________________________________________________________________________
Item No 15:
DIGITAL MASTER PRESSURE GAUGE.
1.RANGE:
2.OVER RANGE PROTECTION 130% OF FULL SCALE
3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING
4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER
5.LCD WITH BACKLIGHT DISPLAY
6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.
7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.
8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION
9.OPERATING TEMPERATURE:
10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.
11.SENSOR MATERIAL:
12.WELDED PARTS: SS316
13.ENCLOSURE: IP 65 PROTECTION CLASS.
14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.
15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.
16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.
17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.
THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.
________________________________________________________________________
Item No 16:
DIGITAL MASTER PRESSURE GAUGE.
1.RANGE: 0 TO 50 BAR
2.OVER RANGE PROTECTION 130% OF FULL SCALE
3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING
4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER
5.LCD WITH BACKLIGHT DISPLAY
6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.
7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.
8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION9. OPERATING TEMPERATURE:
10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.
11.SENSOR MATERIAL:
12.WELDED PARTS: SS316
13.ENCLOSURE: IP 65 PROTECTION CLASS.
14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.
15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.
16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.
17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.
THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.
________________________________________________________________________
Item No 17:
DIGITAL MASTER PRESSURE GAUGE.
1.RANGE: 0 TO 200 BAR
2.OVER RANGE PROTECTION 130% OF FULL SCALE
3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING
4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER
5.LCD WITH BACKLIGHT DISPLAY
6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.
7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.
8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION
9.OPERATING TEMPERATURE:
10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION-
ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.
11.SENSOR MATERIAL:
12.WELDED PARTS: SS316
13.ENCLOSURE: IP 65 PROTECTION CLASS.
14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.
15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.
16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.
17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.
THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.
________________________________________________________________________
Item No 18:
DIGITAL MASTER PRESSURE GAUGE.
1.RANGE: 0 TO 4 KSC
2.OVER RANGE PROTECTION 130% OF FULL SCALE
3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING
4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER
5.LCD WITH BACKLIGHT DISPLAY
6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.
7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.
8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION
9.OPERATING TEMPERATURE:
10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.
11.SENSOR MATERIAL:
12.WELDED PARTS: SS316
13.ENCLOSURE: IP 65 PROTECTION CLASS.
14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.
15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.
16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.
17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.
THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.
________________________________________________________________________
Item No 19:
DIGITAL MASTER PRESSURE GAUGE.
1.RANGE:
2.OVER RANGE PROTECTION 130% OF FULL SCALE
3.ACCURACY: 0.05% OF FULL SCALE OR BETTER INCLUDING
4.DISPLAY RESOLUTION: 5 DIGITS OR BETTER
5.LCD WITH BACKLIGHT DISPLAY
6.UNITS: DISPLAY IN VARIOUS ENGINEERING UNITS. THIS SHOULD BE SE- LECTABLE THROUGH KEYPAD; BAR, mBAR, KG/CM2, mmHg, mmH2O ARE ESSEN- TIAL.
7.ZERO FACILTY: GAUGE SHOULD HAVE "ZEROING" FACILTY.
8.PROCESS CONNECTION: 1/2" BSP MALE, BOTTOM ENTRY.ONE EXTRA SUIT- ABLE ADAPTER TO CONVERT 1/2 BSP TO 1/2 NPT CONNECTION
9.OPERATING TEMPERATURE:
10.POWER SUPPLY: BATTERY OPERATED OR RECHARGEABLE BATTERY (WITH ADAPTER AND OTHER ACCESSORIES FOR CHARGING THROUGH 230V AC, 50 Hz SUPPLY). THE ACCURACY AND FUNCTIONING OF GAUGE SHOULD NOT DEPEND ON BATTERY.IF BATTERY IS VERY LOW THEN GAUGE SHOULD STOP FUNCTION- ING AND IT SHOULD GIVE A LOW BATTERY INDICATION WELL BEFORE FULL DISCHARGE.
11.SENSOR MATERIAL:
12.WELDED PARTS: SS316
13.ENCLOSURE: IP 65 PROTECTION CLASS.
14.CALIBRATION CERTIFICATE: CALIBRATION CERTIFICATE TRACEABLE TO ANY NATIONAL/INTERNATIONAL STANDARD WITH VALIDITY OF ATLEAST 09 MONTHS FROM THE DATE OF SUPPLY FROM THE VENDOR'S PREMISE.
15.PARTY SHOULD MENTION ALL THE PARAMETERS AFFECTING THE UNCER- TAINTY OF MEASUREMENT, FOR EXAMPLE, TEMPERATURE EFFECTS, HUMIDITY EFFECTS, ETC.
16.PARTY SHOULD SUPPLY THE GAUGE IN PROTECTIVE CARRYING CASE SUITABLE FOR STORAGE, HANDLING AND TRANSPORTATION.
17.PARTY SHOULD PROVIDE OPERATION & MAINTENANCE MANUAL OF GAUGE ALONG WITH DETAILED CALIBRATION PROCEDURE FOR RECALIBRATION OF GAUGE AT SITE.
THE VENDOR MUST SUBMIT FILLED TECHNICAL COMPLIANCE CHART POINT WISE W.R.T. OUR SPECIFICATIONS IN THEIR OFFER.